| Case and radiation shield
extensions (increase "B" dimension) |
The "B" dimension
refers to the vertical distance from the cold work surface to the
radiation
shield. When larger than standard amounts of work space are needed, both
the
case and radiation shield lengths may be increased. |
| Custom hole pattern on cold
work surface |
To accommodate special
requirements, the standard pattern of tapped holes in the bottom plate
may be
modified. |
| Exterior finish
|
The exterior finish of all
dewars and refrigerators is anodized aluminum. Gold is the standard
color;
however, blue, red, green, black and purple are also available.
|
| High altitude / balloon
modification |
When the case of a dewar must
be operated at temperatures well below the standard range, indium seals
are
substituted for the standard butyl o-rings.
|
| Increased LHe and/or LN2 capacities
|
For each vessel diameter the standard
length nearly equals the diameter. For additional capacities, without
increasing the diameter, the length of the vessel is increased up to a
factor
of two. For larger increases in capacity and hold time it is better to
increase
the diameter. |
| Inverted operation
|
To facilitate direct optical
access to the work space from above, a modification may be added that
allows
the dewar to operate inverted without significant penalty in hold-time.
Large
windows may be accommodated in this configuration. |
| LHe cooled radiation shield |
To provide additional rejection
of incident IR backgrounds within the dewar, a shield may be added which
attaches directly to the cold work surface. Usually this shield is
fitted with
a removable bottom plate similar to the standard shield. |
| Low temperature modification
(LHe only) |
To facilitate pumping the LHe
bath to temperatures below 1.4K the neck tube is modified. This
modification
permits temperatures as low as 1.2K. |
Low temperature modification
(LN2
only)
|
To pump liquid nitrogen below
the triple point, copper webbing is added to the interior of the cryogen
vessel. This increases thermal contact with the solid nitrogen and
allows for
effective cooling.
|
| Rigid support system
|
Standard dewar supports are
designed to provide excellent hold-times. When greater strength is
needed to
meet special mechanical requirements a more rigid support system is used
in
place of the standard system.
|
| Split case and radiation shield
|
To provide direct access to the
work space, the case and radiation shield are modified by insertion of
demountable joints at the proper location.
|